| SPECIFICATIONS | |
| OPERATION | |
| Temperature display | Melting point and melting point range |
| Start temp. range |
(Ambient + 10 °C) to 396 °C, 0 °C minimum set point |
| Stop temp. range | (Start temperature + 4 °C) to 400 °C |
| Temperature resolution | 0.1 °C |
| Ramp rate |
0.1 °C to 20 °C per minute (0.1 °C increments) |
| Heat-up time | ~10 minutes (50 °C to 350 °C) |
| Cool-down time | ~10 minutes (350 °C to 50 °C) |
| Temperature accuracy |
±0.3 °C (up to 100 °C) ±0.5 °C (up to 250 °C) ±0.8 °C (up to 400 °C) |
| Reproducibility | 0.2 °C |
| Temperature sensor | Pt RTD (built-in) |
| Oven control | Closed-loop PID |
| GENERAL | |
| Display | Back-lit, touchscreen LCD (5.8") |
| Printer interface | RS-232 serial port. Supports Epson compatible dot-matrix printers |
| Computer interface |
USB. All instrument functions can be queried and controlled through a high-level command set |
| MeltView software | Windows compatible (USB port) |
| CAPILLARIES | |
| Dimensions | 1.4 mm to 2.0 mm outside dia., 100 mm length |
| Capacity | Up to 3 tubes simultaneously |
| Fill height | 2 mm to 3 mm |
| Power | 90 to 264 VAC, 47 to 63 Hz, 125 W |
| Operating temperature | 0 °C to 40 °C, non-condensing |
| Weight | ±4 kg |
| Dimensions | 19 x 25 x 21.5 (cm) |
PRODUCT INTRODUCTION
OptiMelt provides a fast and accurate means of automatically determining the melting points and melting ranges of chemical substances. Up to three samples can be analyzed simultaneously. A wide observation window, with an illuminated magnification lens, allows users to observe the samples at all times. A small alumunium heating block is kept under tight computer temperature control, providing extremely linear temperature ramping.
PRODUCT FEATURES
For more specification click here